Contact Us for Quotation | Email1:billy.sz.china@gmail.com |
Click to: Products Map | Email2: mewoi@mewoi.com |
Whatsapp: +8618620373879 |
INTRODUCTION
MEWOI-HTY portable blackbody radiation source is mainly used for real-time calibration of infrared imaging thermometer, infrared temperature gun, etc. to eliminate the error of infrared temperature measurement system caused by the complex environment on site, which can effectively improve the measurement accuracy, improve the reliability of infrared temperature measurement system, and reduce the error detection rate. The portable infrared body temperature gun blackbody furnace is the best ideal product for the calibration of infrared ear temperature gun and forehead temperature gun, which can achieve a emissivity of more than 0.99 through the temperature cycle system developed by our company and the new materials developed by our company. |
MAIN FEATURES
1,Small size, the whole machine is only 2.5kg, convenient for field erection. |
2,There is only one cavity in the isothermal blackbody cavity. The cavity radiation is evenly distributed with the spectrum, and the cavity emissivity is above 0.99. |
3,The automatic heating temperature control method is adopted, which is safe and reliable, the heating speed is fast, and the temperature stability is good. |
4,It is easy to operate, small in size, light in weight and easy to carry. It is not only suitable for laboratory calibration, but also for field calibration. |
5,It uses new technology such as double row digital display measurement value and set value parameter setting, with high precision, multi-function and strong anti-interference ability. |
TECHNICAL SPEICIFICATION
Emissvity | ≥0.99 |
Temperature range(indoor temperature≤22℃) | +35℃--45℃ |
Accuracy | Less than ±0.1℃ |
Resolution | 0.01℃ |
Temperature stability | ±(0.1-0.2)℃/10minutes |
Temperature uniformity | Less than ±0.15C° |
Control Method | PID |
Source Area Size | 50mm |
Cavity depth | 50mm |
Dimension | 225mm*225mm*115mm |
Weight | 2.5KG |
Power Supply | 220V AC 50HZ 100W |
Selection Guide
Model No. | Emissvity | Temperature range (indoor temperature≤22℃) | Accuracy | Temperature stability | Temperature uniformity | Source Area |
0.97±0.02 | +5℃--50℃ | Less than ±0.1℃ | ±0.05℃/1hour | Less than ±0.2C° | 70mm*70mm | |
0.97±0.02 | +35℃--45℃ | Less than ±0.1℃ | ±(0.1~0.2)℃/10min | ±0.2℃ | 78mm*78mm | |
≥0.99 | +35℃--50℃ | Less than ±0.1℃ | ±0.1℃/30min | ±0.1C° | 56mm*50mm | |
≥0.99 | +35℃--45℃ | Less than ±0.1℃ | ±(0.1-0.2)℃/10minutes | Less than ±0.15C° | 50mm*50mm | |