| Function | Wireless high voltage electricity testing, High voltage and induction voltage test, H/L voltage current test |
| Power Supply | Receiver: DC6V, alkaline battery LR03X 4PCS; Detector: zinc-manganese dry battery 6F22, 9V |
| Transmission Distance | Straight-line transmission distance about 30M |
Electricity Test Voltage Range
| 380V~220kV
|
|
| Electricity Test Level | 380V, 6.6kV, 10kV, 35kV, 66kV, 110kV, 220kV |
| Voltage to Earth Display Accuracy | ±15%±5dgt |
| Voltage to Earth Display Range | 0.1kV~150kV(Voltage to earth) |
| Line Voltage Display Accuracy | High voltage overhead line ±15%±5dgt (Other applications: ±25%±5dgt) |
| Line Voltage Display Range | 0.1kV~220kV (display line voltage, according to the voltage to earth multiply √3 |
|
| Current Range | 0.1A~6000A |
|
| Current Accuracy | ±3%±5dgt |
|
| Current Resolution | 0.1A |
|
| Hook Caliber | φ60mm |
| Wireless Frequency | 433Hz |
| LCD Size | 47mm×28.5mm |
| Power Supply Indication | With green power indicator |
| Electricity Test Indication | In electricity test, the detector with sound and light indication function, red double flash indicator and "beep - beep - beep" buzzing sound. |
| Display Speed | 2 Times/Second |
| Data Storage | 99 groups (Power down or replace the battery will not lose data) |
| LCD Backlight | Yes |
| Automatic Shutdown | 15 minutes after boot up, the meter shuts down automatically without any operation |
| Low Battery Voltage Indication | When the detector battery voltage is lower than 7.2V±0.1V, the green power indicator flashes; when the receiver voltage lower than 4.8V±0.1V, low battery voltage symbol will display and remind to replace battery. |
| Rated Current | Detector: 75mA max; Receiver: 35mA max |
| Communication Interface | USB |
| Instrument Dimension | Receiver:78mm×165mm×42mm Detector: 300mm×237mm×85mm |
| Instrument Weight | Instrument: 660g; Package and insulation rod total weight: 5.6kg |
| Insulation Rod Length | Max Diameter:Φ38 mm; Length: Contraction state 850mm; Stretched state 3600mm |
| Insulation Test | The insulation rod stretched to both ends: AC 220 kV / rms |
| Host, detector: AC3700V / rms (between exposed metal and plastic shell) |
| External Interference | Non extra strong electromagnetic field;without 433MHz, 315MHz same frequency interference |
| Working Temperature | -10℃~40℃;below 80%rh |
| Store Temperature | -10℃~60℃;below 70%rh
|